5 Operation | 5.1 Attaching the medical device

5 Operation

Note

At the beginning of each workday, the water-conducting systems should be
rinsed for at least 2 minutes (without transmission handpieces being at-
tached) and if there is a risk of contamination from reflux or back suction, the
system may also need to be rinsed for 20-30 seconds after each patient.

5.1 Attaching the medical device

 WARNING

Detachment of the medical device during treatment.

A medical device that is not properly locked can release from the hose during
treatment.

Before each use, check if the medical device is securely locked onto the
hose by pulling on it.

Connect the coupling part of the medical device to the hose proceeding in
clockwise direction.

Pull on it to make sure that the medical device is securely attached to the
hose.

Note

The water quantity can be regulated through the coupling or the treatment
centre.

5.2 Removing the medical device

Firmly grasp the coupling and pull the medical device off while twisting
slightly.

5.3 Inserting the bur or diamond

Note

Only use carbide burs or diamonds that comply with ISO 1797-1 type 3, are
made of steel or hard metal and meet the following criteria:
- Shaft diameter: 1.59 to 1.60 mm
- Blade diameter: max. 2 mm
SMARTtorque S608 C and SMARTtorque S609 C:
- Overall length: max. 25 mm
- Shaft clamping length: min. 11 mm
SMARTtorque Mini S605 C:
- Overall length: max. 21 mm
- Shaft clamping length: min. 9 mm

 WARNING

Use recommended dental burs or diamonds only.

Risk of injury.

Comply with the instructions for use and the intended use of the bur.

13 / 26

Summary of Contents for SMARTtorque S609 C

Page 1: ...Instructions for use SMARTtorque S609 C 1 008 1644 SMARTtorque Mini S605 C 1 008 1645 SMARTtorque S608 C 3 001 6560...

Page 2: ...stributed by KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Phone 847 550 6800 Fax 847 550 6825 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo c...

Page 3: ...oting 16 6 1 Cleaning the spray nozzle 16 7 Processing steps in accordance with ISO 17664 17 7 1 Preparations at the site of use 17 7 2 Manual processing 17 7 2 1 Manual external cleaning 17 7 2 2 Man...

Page 4: ...airs please contact KaVo Repair Service For scheduling or if you have any questions please contact KaVo Repair Service KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Toll fre...

Page 5: ...tion and storage conditions Air pressure Transportation and storage conditions Humidity range Protect from moisture Keep dry Fragile handle with care HIBC Code Hazard levels The warning and safety not...

Page 6: ...ct and accessories appropriately before disposal In the case of injury to soft tissue do not continue treatment in the oral cavity with instruments driven by compressed air 2 2 Technical condition A d...

Page 7: ...to the device unless these have been ap proved by the manufacturer of the product Use original KaVo spare parts only 2 4 Qualification of personnel Application of the product by users lacking appropr...

Page 8: ...ovided with a fiber optic light system The devices can be sterilized in a steam sterilizer autoclave Through the tube and the MULTIflex coupling con nected to a dental unit the SMARTtorque handpieces...

Page 9: ...C Water consumption 50 ml min Idle speed 350 000 to 400 000 min 1 350 000 to 450 000 min 1 350 000 to 400 000 min 1 Power torque 16 W 0 14 Ncm 13 W 0 12 Ncm 16 W 0 14 Ncm Head size Default Mini Defaul...

Page 10: ...e Mini S605 C 1 008 1645 SMARTtorque S608 C 3 001 6560 3 Product description 3 3 Transportation and storage conditions 10 26 Relative humidity 5 RH to 95 RH absence of condensation Air pressure 700 hP...

Page 11: ...process and sterilize the product and accessories appropri ately See also 2 7 Processing steps in accordance with ISO 17664 Page 17 4 1 Checking the water quantity CAUTION Overheating of the tooth due...

Page 12: ...gauge Mat no 0 411 8731 between hose and medical device and check the following pressures Drive pressure drive pressure recommended return air pressure spray water pressure and spray air pressure SMA...

Page 13: ...ling on it Connect the coupling part of the medical device to the hose proceeding in clockwise direction Pull on it to make sure that the medical device is securely attached to the hose Note The water...

Page 14: ...ement of the tool Property damage to tool shaft and chuck system reduction of the service life of tool and chuck system Do not operate the tool at a higher speed than recommended by the manu facturer...

Page 15: ...008 1644 SMARTtorque Mini S605 C 1 008 1645 SMARTtorque S608 C 3 001 6560 5 Operation 5 4 Removing the bur or diamond After the bur has stopped rotating firmly press the push button with your thumb a...

Page 16: ...ing the spray nozzle WARNING Hazard from non sterile products Infection hazard for dentist and patient Reprocess and sterilize the medical device properly before the next use CAUTION Hazard from insuf...

Page 17: ...tool Remove the tool Process the medical device as soon as possible after treatment To minimize the risk of infection during processing always wear protective gloves Remove the tool from the medical...

Page 18: ...d by the disinfectant manufacturer and proven by an expert opinion Approved disinfectants CaviWipes and CaviCide made by Metrex Consumables required Cloths for wiping the medical device Spray the disi...

Page 19: ...nds washer disinfectors in accordance with ISO 15883 1 which are operated using alkaline cleaning agents For program settings as well as cleansers and disinfectants to be used please refer to the Inst...

Page 20: ...ly Remove the tool from the medical device Position the tip of the spray nipple in the opening and apply the spray Press the spray key once for 1 2 seconds 7 4 2 Servicing with KaVo QUATTROcare PLUS S...

Page 21: ...e minutes with no service procedure running the device automatically switches back to normal service mode See also 2 Servicing with KaVo QUATTROcare PLUS 7 5 Packaging Note The sterile goods package m...

Page 22: ...135 C 275 F Drying time 16 min Sterilizer using the gravity method at least 10 minutes at 135 C 275 F drying time 30 minutes or alternatively at least 60 minutes at 121 C 250 F Drying time 15 min Remo...

Page 23: ...o Spare turbine Standard 1 008 4163 Spare turbine Mini 1 008 4164 Wrench Standard 0 411 3053 Wrench Mini new 1 008 6133 Wrench Mini old 1 006 3384 Nozzle pin 0 410 0921 INTRA spray head incl Midwest a...