Instructions for use SMARTtorque S609 C - 1.008.1644 SMARTtorque Mini S605 C - 1.008.1645 SMARTtorque
S608 C - 3.001.6560
4 Startup and shut-down | 4.1 Checking the water quantity
4 Startup and shut-down
WARNING
Hazard from non-sterile products.
Infection hazard for dentist and patient.
▶
Before first use and after each use, process and sterilize the medical device
and accessories accordingly.
See also:
2
7 Processing steps in accordance with ISO 17664, Page 17
WARNING
Dispose of the product in the appropriate manner.
Infection hazard.
▶
Before disposal, process and sterilize the product and accessories appropri-
ately.
See also:
2
7 Processing steps in accordance with ISO 17664, Page 17
4.1 Checking the water quantity
CAUTION
Overheating of the tooth due to insufficient amount of cooling water.
Insufficient spray water can cause the medical device to overheat and damage
the pulp and tooth.
▶
Adjust the water amount for the spray cooling to a minimum of 50 ml/min
(3.1 inch
3
).
▶
Check the spray water channels and clean the spray nozzles with the nozzle
needle
Mat. no. 0.410.0921
, if needed.
4.2 Checking the pressures
NOTICE
Contaminated or moist compressed air at the compressed air connec-
tion.
Premature wear.
▶
Supply dry, clean and uncontaminated compressed air in accordance with
ISO 7494-2 only.
11 / 26