Page 1: ...Instructions for use SMARTmatic S10 REF 1 011 6730 SMARTmatic S10 K REF 1 011 6731 SMARTmatic S10 S REF 1 011 6732...
Page 2: ...Distributed by KaVo Dental Corporation 11729 Fruehauf Drive Charlotte NC 28273 USA Phone 847 550 6800 Fax 847 550 6825 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...
Page 3: ...s 16 6 2 Troubleshooting 16 6 2 1 Replacing the O rings 16 6 2 2 Cleaning the spray tube 16 7 Processing steps in accordance with ISO 17664 18 7 1 Preparations at the site of use 18 7 2 Non fixing pre...
Page 4: ...Tools and consumables 26 9 Terms and conditions of warranty 27 Instructions for use SMARTmatic S10 REF 1 011 6730 SMARTmatic S10 K REF 1 011 6731 SMARTmatic S10 S REF 1 011 6732 Table of contents 4 3...
Page 5: ...usa com Please direct all questions regarding the product service and maintenance to the KaVo Technical Service Toll free 1 888 ASK KAVO 888 275 5286 Email customerservice kavo com Please refer to the...
Page 6: ...tions Air pressure Transportation and storage conditions Humidity limitations Protect from moisture Keep dry Fragile handle with care HIBC Code Hazard levels The warning and safety notes in this docum...
Page 7: ...d In the case of injury to soft tissue do not continue treatment in the oral cavity with instruments driven by compressed air To test use and remove the tool use a glove or finger guard 2 2 Technical...
Page 8: ...cation of the product by users lacking appropriate medical training can injure the patient the user or third parties Make sure that the user has read and comprehends the instructions for use Only empl...
Page 9: ...le and ergonomically shaped The devices can be sterilized by the steam autoclave method Through the tube and the electrical motor connected to a dental unit the SMARTmatic handpieces equipped with a h...
Page 10: ...plication described herein You need to comply with the following the applicable health and safety regulations the applicable accident prevention regulations these Instructions for use In accordance wi...
Page 11: ...spray clip and the taper sleeve on the dental instrument and or the motor with external spray outlet 3 3 Transportation and storage conditions NOTICE Startup after refrigerated storage Malfunction Pr...
Page 12: ...priately NOTICE Damage from contaminated and moist cooling air Contaminated and moist cooling air can cause malfunctions Make sure that the supplied cooling air is dry clean and free of contamination...
Page 13: ...e to the driver Never attach or remove the handpiece while the drive motor is rotating Lightly spray O rings on motor coupling with KaVo Spray Attach the handpiece to the motor coupling until it locks...
Page 14: ...huck due to excessive speed of the tool or abrupt engage ment of the tool Property damage to tool shaft and chuck system reduction of the service life of tool and chuck system Do not operate the tool...
Page 15: ...he taper sleeve back into its initial position 5 5 Modification for contra angle burs Note Using contra angle burs the handpiece must be converted Open the handpiece chuck Insert the enclosed bur stop...
Page 16: ...or is uneven Service the medical device Missing O ring on the motor coupling Replace O ring See also 2 Instructions for use of motor 6 2 Troubleshooting 6 2 1 Replacing the O rings CAUTION Hazard due...
Page 17: ...EF 1 011 6730 SMARTmatic S10 K REF 1 011 6731 SMARTmatic S10 S REF 1 011 6732 6 Checking for malfunctions and troubleshooting 6 2 Troubleshooting Use the nozzle needle Mat no 0 410 0931 to free the wa...
Page 18: ...s Remove the tool from the medical device Take the spray clip off Remove all residual cement composite or blood immediately Process the medical device as soon as possible after treatment The medical d...
Page 19: ...iminary cleaning is a central constituent and must be carried out prior to the automated processing In the cleaning and disinfecting device validated internal cleaning of the spray tube necessitates p...
Page 20: ...terior of this product is not to be cleaned manually 7 3 4 Automated internal cleaning KaVo recommends washer disinfectors in compliance with ISO 15883 1 which are operated using alkaline cleaning age...
Page 21: ...n Not applicable This product is suitable for automated disinfection only 7 4 3 Automated external and internal disinfection For program settings as well as cleansers and disinfectants to be used plea...
Page 22: ...ress the spray button for one second Servicing the chucking system KaVo recommends cleaning and servicing the chuck system once a week Remove tool place the spray nipple tip in the opening and spray C...
Page 23: ...fore each sterilization Remove the bur Service the product in the QUATTROcare PLUS See also 2 Instructions for use KaVo QUATTROcare PLUS Servicing the chuck KaVo recommends cleaning and servicing the...
Page 24: ...or the sterilization procedure Seal each medical device individually in a sterilization item package 7 8 Sterilization Sterilization in a steam sterilizer autoclave in accordance with ISO 17665 1 NOTI...
Page 25: ...MARTmatic S10 S REF 1 011 6732 7 Processing steps in accordance with ISO 17664 7 9 Storage 7 9 Storage Processed products must be stored protected from bacteria to the extent possible and dust in a dr...
Page 26: ...t No Instrument stand 2151 0 411 9501 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 Nozzle needle 0 410 0931 Spray hose sterilizable 0 065 5188 Bur stop 0 524 0892 Hook 0 410 1963 Sp...
Page 27: ...shall not be liable for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connectio...
Page 28: ......
Page 29: ......
Page 30: ...1 010 0023 s 20170220 2 en US...