9 Technical specifications
52
KaVo Scan eXam One
9.3 Imaging plate specifications
* High resolution mode image sizes are
approximately half of the values in the table.
Imaging plates
Plate size
Size 0
Size 1
Size 2
Size 3
Size 4C
Dimensions (mm)
22 x 31
24 x 40
31 x 41
27 x 54
48 x 54
nominal
Image size (pixels)
*
734 x
1034
800 x
1334
1034 x
1368
900 x
1800
1600 x
1800
nominal
Image size (MB) *
1.44
2.03
2.69
3.09
5.49
nominal
Environ-
mental
condi-
tions
Storage
and
transpor-
tation
-10°C … +40 °C / max 80% RH / NO UV radiation
Use
+10°C …+40 °C / max 80% RH / NO UV radiation
Material
Layer of small photo-stimulable particles (that exhibit the
phenomenom of phosphorescence) uniformly coated on
a support plastic material. Shielded with a protective top
coat layer on the sensitive surface and encapsulated with
lacquer around the edges. Imaging plates do not not in-
clude phosphorous / phosphorus (P).
Use
The typical service life for an imaging plate is several
hundreds of cycles provided that the imaging plate is
handled with care and according to the supplied instruc-
tions. The use of genuine hygiene accessories (protective
covers and hygiene bags) will extend the service life of
the imaging plates.
Disposal
Imaging plates are industrial waste and must be dis-
posed of in accordance with local and national regula-
tions concerning the disposal of such material. Never use
damaged imaging plates.
Approved:
Reviewed: Jalkanen Joni 2017-02-13 16:00
Reviewed
See PDM system to determine the status of this document. Printed out: 2017-03-17 10:35:17
D510414, 5
Copyright © by PaloDEx Group Oy. All rights reserved.