6 Installation of the imaging plate system
KaVo Scan eXam One
35
6 Installation of the imaging plate
system
Imaging plate system is formed of one or more PC
that connects the imaging plate scanner unit,
software, accessories and consumables.
Electronic equipment that does not fulfill medical
safety standards (office PC, network connecting
units etc.) must not be installed in the patient area.
The definition of the patient area is 1,5 m in
horizontal distance and 2.5 m in vertical distance to
the patient. The Scan eXam™ One meets safety
requirements of a medical electrical unit and it can
be installed also in the patient area.
6.1 Positioning the unit
Position the unit on a stable flat surface so that any
potential vibrations do not degrade the image
quality. The unit must not be positioned so that it is
touching other equipment. It must not be placed on
top of or under other equipment.
Do not position the unit in direct sunlight or near
bright light. Sunlight or bright light must not be
allowed to shine directly on the unit door into which
the imaging plates are inserted.
Typical location for a scanner unit in a shared use
by multiple operators is somewhere in a common
space for all users having an easy access to.