Instructions for use PROPHYflex 4 Perio Kit - 1.011.9403 Refill PROPHYflex perio tip - 1.010.0287
6 Reprocessing steps in accordance with ISO 17664 | 6.1 Preparations at the site of use
6 Reprocessing steps in accordance with ISO 17664
WARNING
Hazard from non-sterile products.
Infection hazard for dentist and patient.
▶
Before first use and after each use, prepare and sterilise the medical device
and accessories appropriately.
Note
The reprocessing procedures described in the following apply to the PROPHY-
flex perio tip, adaptor and wrench.
6.1 Preparations at the site of use
WARNING
Hazard from contaminated products.
Contaminated products are associated with an infection hazard.
▶
Take suitable personal protective measures.
Note
Do not place the perio tip in the drill bit bath since the fine capillaries will not
be able to be rinsed under running water any longer and strong corrosion oc-
curs.
▶
Use the wrench to remove the PROPHYflex perio tip from the adaptor.
See also:
2
4.4 Removing the PROPHYflex perio tip, Page 11
Note
Hardened or damaged adaptor O-rings need to be replaced.
▶
Reprocess the medical device as soon as possible after treatment.
▶
The medical device must be dry when transported to reprocessing.
▶
To minimise the risk of infection during reprocessing, always wear protect-
ive gloves.
▶
Remove all residual cement, composite or blood immediately.
▶
Do not place in solutions or similar substance.
6.2 Non-fixing pre-cleaning of the PROPHYflex perio tip
Accessories required:
▪ Demineralised water 30 °C ± 2 °C (86
o
F +/- 4
o
F)
▪ Nozzle pin
▪ Brush, e.g. medium-hard toothbrush
▪ Disposable syringe
▶
Brush the PROPHYflex perio tip with a medium-hard toothbrush under de-
mineralised water.
▶
Check the patency of the PROPHYflex perio tip and if necessary clean it with
the nozzle pin (
Mat. no. 0.410.0911
).
▶
Rinse the PROPHYflex perio tip with at least 20 ml demineralised water us-
ing a disposable syringe.
15 / 24