Instructions for use PROPHYflex 4 Perio Kit - 1.011.9403 Refill PROPHYflex perio tip - 1.010.0287
5 Operation
5 Operation
Preparation
CAUTION
Risk of increased development of emphysema.
If the pressure is too high during the treatment, there is an increased risk of
emphysema developing.
▶
Do not select the highest level for the treatment.
▶
The powder quantity can be controlled at three levels using the adjusting
ring on the PROPHYflex 4.
ð
Select the middle level for the treatment.
Add the powder
▶
Shake the bottle before adding powder.
▶
Add PROPHYflex Perio Powder to the marked level of the powder container.
Add the powder slowly to avoid the formation of dust.
▶
Screw the container filled with the PROPHYflex Perio Powder onto the PRO-
PHYflex 4.
Settings on the powder jet unit
▶
Aim the jet nozzle at a wet washbasin from a distance of approx. 20 cm.
▶
Set the quantities of water and air/powder for treatment as described in the
Instructions for Use of the unit. Never use the powder jet device without
water since this would render the mixture of air/powder more difficult to as-
pirate.
▶
Direct the jet into a washbasin until a homogeneous powder/water mixture
is produced.
Getting the patient ready
▶
Apply some Vaseline
®
ointment to the patient's lips. This prevents the lips
from becoming dry and cracking.
▶
Hook the small saliva ejector in the corner of the patient's mouth such that
the saliva is extracted from under the tongue. Use a large suction cannula
for aspiration of the water/powder mixture bouncing off the tooth.
13 / 24