Instructions for use KaVo polymerisation light Poly One
7 Troubleshooting
22 / 26
7 Troubleshooting
Note
If disturbance arises, a technician authorised by KaVo must be engaged for
the remedial action.
Malfunction
Cause
Remedy
Nothing works
See:
Instructions for
use of the treatment
centre
▶
See:
Instructions for use of the treatment
centre
Handpiece is non-func-
tional
Handpiece is defective.
Supply cable is defect-
ive.
Supply is interrupted.
LED defective
▶
Stop working and call the service technician
Operating button is
non-functional
Handpiece is defective.
Button is defective
▶
Stop working and call the service technician
No timer signal
Handpiece is defective.
Trigger is defective
▶
Stop working and call the service technician
Fibre-optic cable is at-
tached loosely
Socket on fibre-optic
cable is defective.
▶
Replace fibre-optic cable.
O-rings in handpiece
are defective
▶
Replace O-rings
Eye protection is
wobbly
Eye protection is de-
fective.
Rubber part on the eye
protection is defective
▶
Replace the eye protection