Instructions for use KaVo polymerisation light Poly One
3 Description of the Product | 3.1 Technical Specifications
14 / 26
3.1 Technical Specifications
Light source
LED
Sterilisable parts
Fibre-optic cable made of glass
Wavelengths
450 - 460 nm
Peak wavelength
457 nm
Curing times
3 curing times: 20, 40 or 60 seconds
Minimum radiation power
800 mW/cm
2
The radiation power was determined
with a spectroradiometer.
Weight of handpiece and fibre-optic
cable
160 g
Operating temperature
42 °C (at an ambient temperature of
20 °C)
Max. power of the LED
5 W
Max. current of the LED
700 mA
Rated power of the LED
3.3 W
Rated current of the LED
530 mA
Mode: intermittent operation
Operating time
1 minute
Pause time
1 minute
Dimensions
Length including fibre-optic cable
200 mm
Diameter
25 mm
Cross-section of the optically active
surface of the fibre-optic cable
0.385 cm
2
(38.5 mm
2
)
Operating environment
WARNING
Inappropriate operating conditions.
Impairment of the electrical safety of the device.
▶
It is essential to comply with the operating conditions specified in the
"Technical Specifications" chapter.
Ambient temperature
+10 to +40
o
C
Relative Humidity
30% to 75% non-condensing
Air pressure
700 hPa - 1,060 hPa
Max. altitude for operation
up to 3000 m