6 Using the unit
4.
Turn the head dorsal as reference to the canthomeatal line about 35-40° above the
horizontal plane.
5.
Slide the ear rods towards patient’s ears.
6.
Ask the patient open or close mouth.
6.4.2.5 Carpus view (Optional, Not available in USA and
Canada)
CAUTION!
Before taking Carpus image make sure this imaging method is
approved by local authorities of your country.
NOTICE!
If Carpus program button is not displayed in Cephalometric imaging
modality tab, ask local distributor to activate the button. (Not in USA or Canada).
1.
Unlock the lever and turn the ear rods to the PA projection position. Lock the
position. Tilt the nasion support aside. Place the carpus holder to the nasion support
holder.
2.
Adjust the unit height if needed.
3.
Ask the patient to remove rings and metal objects and to place hand on the carpus
holder.
6.4.3 Taking a cephalometric exposure
1.
Protect yourself from radiation by standing behind a suitable x-ray radiation shield.
Make sure that you can see and hear the patient during the exposure.
ORTHOPANTOMOGRAPH
™
OP 3D Pro