6 Using the unit
7.
Adjust the position of the TMJ light until it aligns in the middle of condyle.
NOTICE!
The condyle moves forward by approximately 10 mm when the
mouth is opened.
TMJ light adjustment buttons:
● A) Forwards (towards the mirror)
● B) Reset
● C) Backwards (away from the mirror)
6.3.5 Maxillary Sinus patient positioning
1.
Insert the required positioning devices, bite fork with the bite block on the sinus rest.
Place the disposable covers.
NOTICE!
Use a new disposable cover for every patient.
2.
Adjust the unit height.
3.
Guide the patient to the unit and instruct to stand as straight and tall as possible.
Ask the patient to take grip on the handles and bite on the bite block.
4.
Adjust the height of the Frankfort-Horizontal plane (FH) light so that it goes through
the patients Porion and Orbitale. Straighten the patient's head if needed.
ORTHOPANTOMOGRAPH
™
OP 3D Pro
55