4 Imaging programs
Pediatric Lateral Cephalostat has an optimized image
height (180 mm) that is used e.g. for pediatric patients
but also adult patients to reduce the radiation dose.
The pediatric lateral projection covers all the typical
cephalostat landmarks from nasion down to the spine.
The starting point of the lateral scan is adjustable.
Exposure settings for cephalometric pediatric lateral program
100 VAC 90 kV/ 8 mA/ 10
s/ 29 mGycm
2
90 kV/ 8 mA/ 13
s/ 36 mGycm
2
90 kV/ 8 mA/ 16
s/ 46 mGycm
2
90 kV/ 8 mA/ 20
s/ 57 mGycm
2
120 VAC 85 kV/ 10 mA/ 10
s/32 mGycm
2
85 kV/ 10 mA/ 13
s/40 mGycm
2
90 kV/ 8 mA/ 16
s/ 46 mGycm
2
90 kV/ 8 mA/ 20
s/ 57 mGycm
2
230 VAC 85 kV/ 10 mA/ 13
s/40 mGycm
2
90 kV/ 10 mA/ 13
s/45 mGycm
2
90 kV/ 12.5 mA/
16 s/72 mGycm
2
90 kV/ 12.5 mA/
20 s/90 mGycm
2
Cephalometric lateral projection
Lateral Cephalostat uses full height image field that
provides a nearly full skull image. The starting point of
the lateral scan is adjustable.
ORTHOPANTOMOGRAPH
™
OP 3D Pro