Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640
7 Reprocessing steps in accordance with ISO 17664 | 7.2 Manual Reprocessing
21 / 28
7.2.3 Manual external disinfection
WARNING
Incomplete disinfection.
Infection hazard.
4
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
4
If the disinfectants/disinfection procedures fail to meet the mandatory na-
tional requirements, perform a final sterilisation with the sterilisation para-
meters as described.
CAUTION
Never reprocess the medical device with chloride-containing products.
Malfunction and material damage.
4
Reprocess in a washer disinfector or by hand only.
KaVo recommends the following products based on the compatibility of the ma-
terials. The microbiological efficacy must be ensured by the disinfectant manu-
facturer and proven by an expert opinion.
Approved disinfectants:
▪ CaviWipes and CaviCide made by Metrex
Consumables required:
▪ Cloths for wiping the medical device.
4
Spray the disinfectant on a cloth, then wipe down the medical device and al-
low the disinfectant to act according to the instructions of the disinfectant
manufacturer.
4
Follow the instructions for use of the disinfectant.
7.2.4 Manual internal disinfection
This product is not designed for manual internal disinfection. For effective re-
processing, the inside of the device must be subjected to automated cleaning in
a cleaning and disinfection unit in accordance with ISO 15883-1.
7.2.5 Manual drying
4
Blow off the outside and inside with compressed air until water drops are no
longer visible.
4
Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.
Also refer to: