7 Reprocessing steps in accordance with ISO 17664 | 7.3 Automated reprocessing
22 / 28
7.3 Automated reprocessing
WARNING
Incomplete disinfection.
Infection hazard.
4
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
4
If the disinfectants/disinfection procedures fail to meet the mandatory na-
tional requirements, perform a final sterilisation with the sterilisation para-
meters as described.
WARNING
Sharp dental bur in the medical device.
Injury hazard from sharp and/or pointed dental bur.
4
Remove the dental bur.
CAUTION
Never reprocess the medical device with chloride-containing products.
Malfunction and material damage.
4
Reprocess in a washer disinfector or by hand only.
CAUTION
Never reprocess this medical device in an ultrasonic cleaner.
Malfunction and material damage.
4
Reprocess in a washer disinfector or by hand only.
7.3.1 Automated internal and external cleaning and
internal and external disinfection
KaVo recommends washer disinfectors in accordance with ISO 15883-1, which
are operated using alkaline cleaning agents.
4
For programme settings as well as cleansers and disinfectants to be used,
please refer to the Instructions for Use of the washer disinfector.
7.3.2 Automated drying
The drying procedure is usually part of the cleaning programme of the washer
disinfector.
NOTE
Please comply with the instructions for use of the washer disinfector.
4
In order not to affect the KaVo medical device, make sure that the product
is dry on the inside and outside after completion of the cycle.
4
Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.