7 Reprocessing steps in accordance with ISO 17664 | 7.5 Packaging
24 / 28
NOTE
Handpieces must be taken off the service couplings before the chuck service
can be started and performed.
4
Close the front flap and press the chuck service button for at least three
seconds until the spray canister control LED flashes three times consecut-
ively.
ð
The device is in chuck service mode.
4
Remove the service coupling of the chuck from the side hatch of the
QUATTROcare PLUS and attach it to coupling service point four, on the far
right. A MULTIflex adaptor must be mounted there.
4
Press the handpiece together with the guide bush of the chuck to be ser-
viced against the tip of the service coupling.
4
Press the button marked with the chuck service symbol.
NOTE
Close the chuck service mode.
Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A,
close the front flap and start the service procedure.
Option 2: After three minutes with no service procedure running, the device
automatically switches back to normal service mode.
Also refer to:
7.4.2 Servicing with KaVo QUATTROcare PLUS, Page 23
7.5 Packaging
NOTE
The sterile goods package must be large enough to accommodate the product
without stretching the packaging. The quality and use of the packaging of the
items to be sterilised must meet the applicable standards and be appropriate
for the sterilisation process!
4
Seal the medical device separately in a sterile pack.
7.6 Sterilisation
Sterilisation in a steam steriliser (autoclave) in
accordance with ISO 17665-1
Sterilisation has to be performed using a steam steriliser (autoclave) approved
by the U.S. Food and Drug Administration (FDA).
CAUTION
Improper service and care.
Risk of injury.
4
Service regularly with suitable agents.
CAUTION
Contact corrosion due to moisture.
Damage to product.
4
Immediately remove the product from the steam steriliser after the steril-
isation cycle.