Instructions for use EXPERTsurg LUX – 1.008.3500
2 Safety | 2.6 Improper use
9 / 68
2.6 Improper use
Parts that rotate when the pump is running may injure the patient, user and
third parties.
4
Do not reach into the pump. Turn off the device when the pump is open.
2.7 Accessories and combination with other equipment
Use of un-authorized accessories on the device or un-authorized modifications
to the device can lead to injury.
4
Only use accessories that have been approved for combination with the
product by the manufacturer.
4
Do not make any modifications to the device unless these have been ap-
proved by the manufacturer of the product.
Improper use of handpieces can cause injuries.
You need to comply with the following guidelines to ensure the safe use of the
electrically-driven handpieces:
4
Comply with the instructions for use of the respective handpiece.
4
Check the speed setting each time you turn on the device.
4
Comply with the permissible maximum speed and maximum contact pres-
sure of the burs as specified by the bur manufacturer.
4
Comply with the servicing instructions for handpieces as specified in the re-
spective instructions for use.
The bluish LED light of the motors can damage the cornea or lens of the eye.
4
Do not gaze into the lamp when it is in operation.
4
Use adequate shielding for eye protection.
2.8 Qualification of personnel
Application of the product by users lacking appropriate medical training can in-
jure the patient, the user or third parties.
4
Make sure that the user has read and comprehends the instructions for use.
4
Only employ the device if the user has the appropriate medical training.
4
Comply with national and regional regulations.
2.9 Service and repair
Servicing work that is described in the "Servicing" chapter of the present In-
structions for Use can be performed by the operator/user.
Repairs and safety checks may only be performed by trained service personnel.
The following persons are authorized to do this:
▪ Service technicians of KaVo branches after the appropriate product training
▪ Service technicians of KaVo authorized dealers after the appropriate product
training
Comply with the following items during all servicing work:
4
Have the service and testing tasks carried out in accordance with the Medi-
cal Device Operation Ordinance.
4
After interventions on and repairs of the device and before re-use, have the
service personnel perform a safety check on the device.
4
The device should be serviced, cleaned, and stored in a dry location, and
should be disconnected from the power system, according to instructions, if
it is not to be used for extended periods.