Instructions for use EXPERTsurg LUX – 1.008.3500
5 Operation | 5.8 Foot control
46 / 68
5.8 Foot control
5.8.1 Changing the speed, coolant flow rate, and
direction of motor rotation
Control element
Function
4
Press the foot pedal of the foot control in order to
start the motor and increase the speed.
4
Depending on the pump key operating mode of the
foot control, press the pump key to set the coolant
flow rate or to switch it on or off.
4
Press the pump key of the foot control for an ex-
tended time to activate the rinsing function.
4
Press the direction of motor rotation key of the foot
control to set the direction of motor rotation.
The direction of motor rotation can be changed during
the treatment using the direction of motor rotation key
of the foot control. The changed direction of motor ro-
tation is shown on the display. Counterclockwise rota-
tion is not saved.
A total of 3 audible signals indicate counterclockwise
rotation. When the motor starts up, 3 more audible
signals are issued.
5.8.2 Selecting the program steps
Program steps can be selected during the treatment using the program key of
the foot control.
Control element
Function
4
Briefly press the program key of the foot control to
select the next program step.
ð
After the last program step follows the first step
again.
4
Press the program key of the foot control for a long
time to select the previous program step.
ð
After the last program step follows the first step
again.