Instructions for use EXPERTsurg LUX – 1.008.3500
5 Operation | 5.5 Changing default values
40 / 68
NOTE
The torque values may deviate by max. ± 10 % in the KaVo SURGmatic S201
contra-angle handpieces. Larger deviations are possible with other contra-an-
gle handpieces.
4
Press the plus and minus keys to set the desired value. The step width de-
pends on the transmission ratio and the range of values.
ð
Changed values are stored automatically.
The maximum torque value reached is displayed during the treatment in the
activities "Tapping", "Set implant" and "Set closure cap". The value is re-set as
soon as the motor starts up again.
5.5.3 Setting the coolant flow rate
CAUTION
Coolant dosed incorrectly.
Tissue damage.
4
Please note the instructions for use of the attachment tool.
4
Set a sufficient coolant flow rate.
The coolant flow rate of the KaVo surgical units can be set to 4 levels or
switched off:
Displayed informa-
tion
Description
Off
Level 1 = 32 ml/min
Level 2 = 50 ml/min
Level 3 = 76 ml/min
Level 4 = 110 ml/min