Configuration
B 3.4 SP1 - Storing and recalling motor parameters
All motor-dependent parameters (P41 ... P96) can be stored in separate memories or recalled therefrom, 8
memory locations M1...M8 being available. The corresponding functions are accessible via the menus Special
functions - motor parameters. See diagram in Section B 3.
First select the menu option “Special functions” with key
!
and call up with Enter key
"
. Select the menu entry
“Motor parameters” with the keys
%
and
#
and call up with Enter key
"
. Each special program can be exited
with the Cancel key
!
, the system jumping back in each case to the next highest menu level.
SP11 -
Special functions - motor parameter - recall - memory M1...M8
The selected memory is loaded into the parameters P41 ... P96; these parameters are used for motor control.
Only occupied memories can be recalled. If no memory is occupied, the recall function is not available.
SP12 -
Special functions - motor parameter - store - memory M1...M8
The parameters P41 ... P96 are stored. If a memory is already occupied, you will be asked whether the memory
is to be overwritten (SP121); the memory is overwritten using the Enter key
"
and the function is aborted with-
out storing when the Cancel key
!
is used.
SP13 -
Special functions - motor parameter - delete - memory M1...M8
The memory is deleted. Only occupied memories can be deleted. If no memory is occupied, the delete function
is not available.
SP14 -
Special functions - motor parameter - factory setting
With this menu option, all motor-dependent parameters P41 ... P96 can be set to the factory setting or to various
predefined motors.
SP141 -
Special functions - motor parameter - factory setting - factory default
With this function, motor parameters P41 ... P96 are set to the factory setting. The configuration of the inverter
P1 ... P8 and the configuration of the braking resistance and of the remote control P100 ... P150 remain
unchanged.
SP142 -
Special functions - motor parameter - factory setting- KaVo type xxxx
With this function, the motor parameters P41 ... P96 are preset to values for specific KaVo spnidles. Depending
on use and operating point, the corresponding parameters must be adapted.
B 3.5 SP2 - Test remote control
These functions serve for checking the function of the remote control at X4.
SP21 -
Special functions - test remote ctrl - digital input
The state of the 6 inputs read in by the inverter is displayed, L representing low input voltage (0 V) and H high
input voltage (24 V).
SP22 -
Special functions - test remote ctrl - relay output
With this menu, the switching function of the relays can be checked. First, all 4 relays are switched on. The relay
can be selected (Rx flashes) with the keys
%
and
#
, and the corresponding relay can be switched with the
Enter key
"
. The switching state is displayed as 0 (contact open) and 1 (contact closed).
SP23 -
Special functions - test remote ctrl - analogue input
The voltages read in by the inverter at the inputs AIN1 and AIN2 are displayed.
SP24 -
Special functions - test remote ctrl - analogue output
With this menu, the functioning of the outputs AOUT1 and AOUT2 can be checked.
First, the desired output is selected with the keys
%
and
#
and confirmed with the Enter key
"
. Now, the out-
put voltage can be set with the keys
%
and
#
, confirmation with the Enter key
"
being unnecessary.
29