Instructions for use ERGOcam One
4 Commissioning | 4.2 Carry out a safety check.
CAUTION
Wrong insertion into the USB socket.
USB - plug can only be inserted into the correct position. Forceful insertion leads to
destruction of the plug.
▶ Check the correct position before inserting the plug.
After connecting the camera to the PC, check the correct installation of the camera on
the PC. The camera uses uses system drivers in windows, installation follows auto-
matically by plugging into a USB port.
Note
If the camera is plugged into another USB port of the PC, the installation for this
USB port will be automatically repeated.
Note
In the case of turbulent video operation, an unsuitable front USB port has been
used which is not optimally connected to the main board of the PC. Plug the camera
into an alternative USB port of the PC.
Verification of correct installation:
▪ A further USB input device is displayed under device manager (human interface
device) following connection of the camera.
▪ A device "ERGOcam One" is displayed in the device manager under image pro-
cessing device.
Note
An additional software e.g. Conexio is required for the display / storage of camera
images on the PC.
The assembly of the camera is described in the instructions for assembly and use of
the respective treatment unit.
4.2 Carry out a safety check.
WARNING
Danger of injury from electric current.
Electric shock.
▶ Carry out a safety check after connecting the camera.
▶ Do not continue working with a damaged camera!
▶ Only connect accessories that conform with IEC 60601-1, IEC 60950 or IEC
601010 to the system.
▶ Let the service technician carry out the safety check.
▶ Check the housing and the connection cable of the camera for cracks and dam-
age.
See also:
2
Instructions for use of the treatment unit.
19 / 30