▶ Simultaneously press the air button ① and water button ④ and continuously
increase or decrease the exiting spray by applying more or less pressure on
the two buttons.
5.8.1
Setting heating and cold light using the multifunction hand‐
piece
Afterglow time and cold light intensity are constant.
▶ Remove the multifunction handpiece from the holder.
▶ Press the preselected spray button.
or
▶ Press the preselected spray button on the multifunction foot control.
The LED for the default spray button shines: The heating and cold light for the
multifunction handpiece have been preselected.
▶ To store settings, hold the LP/AP button, with the instrument removed, until
you hear a signal.
5.9
Using Satelec mini LED
See also: Satelec Mini LED Instructions for Use
5.10
Using the spray mist suction device and saliva ejector
Note
If the saliva ejector is used, the slide of the spray mist extraction device must be
closed.
If the spray mist extraction device is used, the slide of the saliva ejector must be
closed.
Suction is automatically activated when the spray mist extraction device or the
saliva ejector is taken out of the holder.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH
5 Operation | 5.9 Using Satelec mini LED
31/48