Technical Specifications
Space needed for DSEclinical type 5197/5197 eH (single chair)
Width
approximately 486 mm
Depth
approximately 250 mm
Maximum height
2300 mm
Weight
Approximately 70 kg
Nominal voltage
Nominal voltage
(country specific)
100 V / 50/60 Hz
120 V / 50/60 Hz
230 V / 50/60 Hz
Mains fuse
Type
T6.3 H / 250 V
Maximum permissible mains voltage
fluctuations
±10 % of the nominal voltage
Overvoltage category
II
Media
Water quality
Tap water
Water hardness
1.5 to 2.14 mmol/l ≙ 8.4 to 12° dH
pH
7.2 to 7.8
Connected loads
Power consumption
approx. 450 VA
Water pressure at customer premises
2 to 6 bar
Air pressure at customer premises
5 to 7 bar
Maximum water consumption
0.1 l/min
Maximum air consumption with external
suction
60 l/min
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH
3 Product description | 3.3 Technical Specifications