Instructions for use Treatment light 9010 for MASTERspace® and ERGOspace®
4 Assembly and start-up | 4.3 Assembling and connecting to the MASTERspace® workplace
4.3 Assembling and connecting to the MASTERspace
®
workplace
4.3.1 Mounting individual and double light stands
Fasten the light stand to the installation support
▶ Clamp the connector ⑦ with two clamping strips ⑧ on the installation support ⑤.
▶ Insert two connecting sleeves ③ in the light stand ③ and align them so that the
slots ① are vertical.
▶ Mount the light stand ⑥ on the connectors ⑦ and affix it with hexagon socket
screws ②.
▶ Mount the plastic covers ⑨.
Fasten the support arm to the light housing without the height
adjustment
▶ Pull off the left cap ② and remove the rear wall ①.
23 / 38