Instructions for use Treatment light 9010 for MASTERspace® and ERGOspace®
4 Assembly and start-up | 4.2 Assembling and connecting to the ERGOspace® workplace
20 / 38
Premounted light stand
▶ Screw two threaded bolts ④ from the fastener set into the baseplate ③ that is
fastened with four screws ② to the light stand ①.
Mount the light stand
▶ Move the premounted light stand ① over the worktop.
▶ Insert the connecting line ⑤ through the middle hole.
▶ Insert four threaded bolts ④ in the 11 mm dia. holes.
▶ Screw four threaded bolts ④ into the bottom of the desktop. Make sure that the
connecting line ⑤ is not is not pinched.
▶ Check if the light stand ① is perpendicular and align if necessary.
▶ Screw the base covers ② into the baseplate ③.