Instructions for use PROPHYflex 3 - 2018 - 1.000.4672, - 1.006.9926, - 1.006.9928
6 Troubleshooting | 6.1 Cleaning a blocked cannula
18 / 26
6 Troubleshooting
Preventive measures
After each treatment and before each sterilisation, unscrew the powder container in an
anticlockwise direction and replace it with a clean powder container. Mount the PRO‐
PHYflex on the MULTIflex (LUX) / MULTIflex LED coupling and blow through the air
and water channels. The powder reservoir should be cleaned with a cloth once per
week.
6.1 Cleaning a blocked cannula
▶ Remove the cannulas.
▶ Twist the nozzle needle from the front and then from the back into the cannula.
▶ The remove the nozzle needle, and blow out the cannula with compressed air.
6.2 Cleaning the main body
To prevent blockages, we recommend cleaning the main body once every week and
prior to each thermal disinfection or sterilization.
▶ Remove the cannulas.
▶ Unscrew the powder container anticlockwise.
▶ Unscrew the P-nozzle in an anti-clockwise direction and push through the aper‐
ture with the nozzle pin.
▶ Clean the media tube with a cleaning drill or remove obstructions.
▶ Then blow through with compressed air.