Instructions for use PROPHYflex 3 - 2018 - 1.000.4672, - 1.006.9926, - 1.006.9928
7 Preparation methods according to ISO 17664 | 7.1 Preparation at the site of use
7 Preparation methods according to ISO 17664
7.1 Preparation at the site of use
WARNING
Hazard from nonsterile products.
There is a risk of infection from contaminated medical devices.
▶ Take suitable personal protective measures.
▶ Immediately remove all residual blood.
▶ Recondition the medical device as soon as possible after treatment.
▶ By unscrewing in an anticlockwise direction, remove the powder container from
the handpiece.
▶ Empty the powder container before preparing the medical device.
▶ All residual powder has to be removed, especially from the cannula, the tube, and
the P-nozzle.
▶ The medical device must be dry when transported for reconditioning.
▶ Do not place it in a solution or similar.
7.2 Cleaning
CAUTION
Malfunctions from cleaning in the ultrasonic unit.
Defects in the product.
▶ Only clean manually or in a thermodisinfector.
7.2.1 Manual cleaning of the exterior
Required accessories:
▪ Tap water 30
o
C ± 5
o
C (86
o
F ± 10
o
F)
▪ Brush such as a medium hard toothbrush
▶ Brush off under flowing tap water.
7.2.2 Automated external cleaning
KaVo recommends thermodisinfectors in accordance with EN ISO 15883-1, which are
operated with alkaline cleaning agents with a pH value of max. 10 (e.g. Miele G
7781/G 7881 – Validation was carried out with Programme "VARIO-TD", cleaning
agent "neodisher® mediclean", neutralisation agent "neodisher® Z" and rinsing agent
"neodisher® mielclear" and only applies to the material compatibility with KaVo prod‐
ucts).
▶ For program settings as well as cleansers and disinfectants to be used, please re‐
fer to the Instructions for Use of the thermodisinfector (complying with max. pH
value of 10).
7.2.3 Manual cleaning of the inside
Can only be done with KaVo CLEANspray or KaVo DRYspray.