K-3 & K-4 Series Owner’s Manual
Page 25
CPR FUNCTION
{Figure 1, Page 15}
Mattress Deflate
Bolsters Deflate
Air Pad Deflate
1.
Air Mattress.
a. To deflate the mattress for a CPR procedure, pull the hose (V) from the control unit connector (press connector
latches in OEM series then pull connector). If the mattress is equipped with bolsters and/or air pad, unplug the
corresponding deflate plug located at the foot end of the mattress.
ELITE SERIES
OEM SERIES
b. In case of CPR emergency and quick deflation of the mattress, unzip the top sheet from the foot to the head by
pulling the zipper located by the patient right foot near the exit location of the hose assembly; or on some
mattresses by unfastening the top sheet straps from the side of the mattress. To disconnect a few air cushions
which are directly below the patient’s chest from the mattress, press the quick release button on the connector
with one hand and pull the air cushion connector with the other.
2.
Foam Aire Mattress.
a.
To deflate the FOAM AIRE mattress for CPR procedure, disconnect the mattress deflate connector which is attached to the
“DEFLATE” tag from the deflate valve. If connected to a control unit, press the quick release buttons on both the coupling
bodies and simultaneously pull the hose from the control unit flange connector
.
PLEASE NOTE: Before using the mattress please make sure that the mattress deflate connector is re-connected back into the
mattress deflate valve.