easily replaced by simply removing and installing a new one. The failed sensor can be sent back to
Kanomax for refurbishment or disposal. Monitor status conditions are as follows:
Fault
Description
LCD Screen
0-5V
Output
(MODEL
2750 only)
Control
Output
(MODEL
2750 only)
Low Alarm
Output
(MODEL
2750 only)
High
Alarm
Output
(MODEL
2750 only)
Data
Logging
(MODEL
2750 only)
No Fault
Valid gas
reading
Reading
proportional
to gas
reading
As set by
user
As set by
user
As set by
user
Valid gad
reading
Sensor Failed
Fault
Sensor failure
5 V
Open
Closed
Closed
9999
Sensor Aging
Fault
Valid gas
r
sensor aging
Reading
proportional
to gas
reading
As set by
user
As set by
user
As set by
user
Valid gad
reading
Communications
Failure
SENSOR
NOT
CONNECTED
or N.C. or
turn off when
use Ni-MH
battery only
0 V
Open
Open
Open
No output
Sensor Standby
Standby
symbol
0 V
Open
Open
Open
No output
8.2. Technical Support
Technical information, service and spare parts are available through your distributor. In addition,
worldwide technical support is available from Kanomax USA, Inc..
Please contact:
Kanomax USA, Inc.
219 US Hwy 206, Andover, NJ 07821
Phone: 973-786-6386
Fax: 973-786-7586
Email: [email protected]
Page | 33
Kanomax Gasmaster User Guide
MRK-D-0002 V3