Note 2: The data log can also be cleared directly from the main display screen by holding
down the mute button for 2 seconds until the monitor beeps (if the monitor keys are not
muted).
•
The display will then give the option to start logging
•
Press the “scroll up” button to select either ON or OFF and press “enter” to confirm the
selection and return to the main menu.
Note 3: Data logging can also be both started and stopped directly from the main display
screen by pressing the “scroll up” button for two seconds until the monitor beeps (if the
monitor keys are not muted).
When data logging is initiated a triangle will appear on the main display screen.
The triangle is an indication that data logging is active, and is also used as an indication of the
presence of data, as described below:
•
An empty triangle indicates that no data is stored
•
A filling triangle indicates that some data is stored in the memory. The triangle will fill up in
10% increments.
•
A full triangle indicates that the memory is full and needs downloading to the PC.
4.
Kanomax Gasmaster Software (Model 2750)
4.1. Set Up
Computer Requirements
•
USB to monitor cable
•
Window OS version 2000 or later
•
Recommended 512 Mb RAM or more
•
Recommended 1 GHz processor speed or faster
CLEAR LOG
YES NO
LOGGING
ON OFF
LOG FREQ.
1 MIN