NanoParticle Nebulizer: Operation
Liquid NanoParticle Sizer User Guide
52
Version 1.0.1
2.
Touch the
Device
button.
3.
On the
AFIMC Device Settings
screen, touch buttons or textboxes to
configure the following:
▪
Charge Conditioner
allows you to change settings of the charge
conditioner.
o
X-Ray
shows current working status of the X-Ray charge
conditioner.
o
Purge
is the
Flow On/Off
status of charge conditioner purge
flow.
o
Setpoint
is the setpoint flow rate of the charge conditioner
purge flow in ccm.
▪
Mobility Classifier
allows you to change settings of the ion mobility
classifier.
o
Voltage
shows the current working status of the ion mobility
classifier high voltage power supply.
o
Control
is the source of control for the high voltage power
supply, confirm it is set to
Program
for LNS operations.
o
Setpoint
is the setpoint voltage for the high voltage power
supply when its control is set at
Direct
.
o
Inlet Flow
is the total aerosol flow rate at the inlet of the
AFIMC, confirm it is set at 1500 ccm for LNS operations.