NanoParticle Nebulizer: Operation
Liquid NanoParticle Sizer User Guide
22
Version 1.0.1
Connecting Gas Port B to Gas Port C
Follow these instructions to connect Gas Port B to Gas Port C:
1.
If you have not already done so, remove the protective shipping caps
from Gas Ports B and C.
2.
Remove the
Gas Port Coupler
from the LNS Accessory Kit.
3.
Connect the
Gas Port Coupler
to Ports B and C. Tighten the Swagelok
nuts on the coupler. Do not tighten more than 1/8 turn past finger tight
Connecting CDA to Gas Port A
Follow these instructions to connect Soft x-ray purge flow gas supply to Gas
Port A:
1.
If you have not already done so, remove the protective shipping caps
from Gas Port A.
2.
Provide a gas supply line of Nitrogen or Compressed Dry Air. Note:
It is highly recommended to add secondary oil and vapor filters to the
gas supply when not using laboratory grade gas. Contact Kanomax
FMT for more information.
3.
Connect a gas supply line using Swagelok nut and ferrule. When first
connecting line, fully swage the nut and ferrule by turning the nut
one full turn past finger tight. Use a second wrench to keep the
bulkhead fitting from rotating. Subsequent connections require only
1/8 turn past finger tight.
Gas Port Coupler