Page 30
MAINTENANCE
EMPTYING AND CLEANING THE INTEGRAL WATER TRAP
The integral water trap should be checked and emptied on a regular
basis, water vapour will condense and gather in the probe line. This may
move suddenly to the trap when the probe is moved, care should be
taken at all times.
EMPTY THE WATER TRAP
Carefully remove the water trap from its housing in the analyser.
The water trap has three parts, carefully split off the bottom part to
access the condensate container. Dispose of the condensate in a
suitable drain – care must be taken as is could be acidic. If condensate
spills onto the skin or clothing, clean off immediately using fresh water
and seek medical advice if problems occur.
CHANGING THE PARTICLE FILTER
This is a very important part of the analyser and should be changed
regularly, it prevents dust and dirty particles entering the pump and
sensors and, hence causing damage. The filter MUST be changed when
it is discoloured.