Page 19
USING THE ANALYSER
EVERY TIME YOU USE THE ANALYSER
BEFORE SWITCH-ON CHECK THAT:
•
The particle filter is dry and clean
•
The water trap and probe line are empty of water
•
All hoses connections, etc, are properly made
•
The probe is sampling CLEAN AMBIENT air
•
The water trap is correctly fitted and instrument upright
•
The flue temperature is connected
Switch ON the instrument by pressing
AUTOMATIC CALIBRATION
During this sequence the analyser pumps fresh air into the sensors to
allow toxic sensors (if fitted) to be set to zero and the oxygen sensor to
be set to 20.95%.
After switch-on the analyser will briefly display header information:-
Software version number and analyser serial number.
Then “ANALYSER PURGING 90 secs” countdown will appear on top of
the main screen.
The calibration time counts down in seconds to zero, calibration time
may be changed to 90, 120, 180, 300 seconds.
NOTE:-
Three minutes is recommended to allow the sensors to stabilise
fully, anything less than this may result in drift of the toxic and oxygen
sensors in clean ambient air.