14. ANALYSER SPECIFICATION
(NOTE MAY BE SUBJECT TO CHANGE)
Parameter
Range
Resolution
Accuracy
Temp Measurement
Flue Temperature
0-600
C
0.1
C
±2°C
±0.3% reading
Inlet Temperature
(Internal sensor)
0-50
C
0.1
C
+1°C
±0.3% reading
Inlet Temperature
(External sensor)
0-600
C
0.1
C
±2°C
±0.3% reading
Flue Gas
Measurement
Oxygen
*2
0.21%
0.1%
+0.3%
Carbon Monoxide
(H
2
compensated)
0-20ppm
21-4,000ppm nom
4,001-10,000ppm
20,000ppm max
for 15 mins
1ppm
+3ppm
+5% of reading
+10% of
reading
Carbon Dioxide
*1
0-20%
0.1%
+0.3% volume
Efficiency (Net or
Gross)
*2
0-99.9%
0.1%
+1.0% reading
Efficiency High (C)
*2
0-119.9%
0.1%
+1.0% reading
Excess Air
*2
0-250%
0.1%
+0.2% reading
CO/CO
2
ratio
*2
0-0.999
0.0001
+5% reading
*1
Using dry gases at STP
*2
Calculated