12. ANALYSER PROBLEM SOLVING
If any problems are not solved with these solutions, contact us or an
authorised repair center.
Fault symptom
Causes / Solutions
Oxygen too high
CO
2
too low
Air leaking into probe, tubing, water
trap, connectors or internal to analyser.
CO reading (- - - -)
Analyser was stored in a cold
environment and is not at normal
working temperature.
CO sensor needs replacing.
Pump is switched off
Batteries not holding
charge
Analyser not running on
mains adapter.
Batteries exhausted.
AC charger not giving correct output.
Fuse blown in charger plug.
Analyser does not
respond to flue gas
Particle filter blocked.
Probe or tubing blocked.
Pump not working or damaged with
contaminants.
Net temperature or
Efficiency calculation
incorrect.
Ambient temperature set wrong during
Automatic Calibration.
Flue temperature
readings erratic
Temperature plug reversed in socket.
Faulty connection or break in cable or
plug.
T flue or T nett displays
-(-OC-)
Probe not connected.
X-Air, EFF display (-O>-)
CO
2
reading is below 2%.
Analyser just continually
beeps
Turn dial back to MENU and press
ENTER