Page
15
4.2 PRESSURE
TEST
Select “Prs”. The pump stops automatically. Press
/
to auto-zero the
pressure sensor. Using the black connectors and manometer hose, connect to P1 for
single pressure or P1 and P2 for differential pressure.
PRS display
PRESSURE
→
Normal response or smoothed (damped)
response can be selected via “Menu”.
P
0.01
→
‘High’ or ‘Low’ resolution readings can be
selected via “Menu”.
mbar
→
Pressure units can be selected via “Menu”.
14:56:29
→
Displays time to enable manually timed tests.
Press
/
to print a full pressure test, (or send to PC via optional
Bluetooth).
Hold
/
for 2+ seconds to log a pressure report.
Viewing / printing a logged pressure test
Select MENU / REPORT / PRESSURE / VIEW
Use
or
to select the log number to be printed.
Press
/
to print the test, (or send to PC via optional Bluetooth).
WARNING
Before using the KANE451 to measure the pressure of a gas/air ratio valve, read the
boiler manufacturer’s instructions thoroughly. If in doubt contact the boiler
manufacturer.
After adjusting a gas/air ratio valve it is essential that the CO, CO
2
and CO/CO
2
ratio
readings are within the boiler manufacturer’s specified limits.