Installation guide flowIQ® 4200

4.3  Split flange installation

Installation sketch:

Installing a meter with split flanges gives the option to install the flanges on the pipeline 
systems first or on the meter itself. Best practice depends on the installation scenarios 
and the space conditions.
Both options are described below and both installation options require the same 
procedure for tightening the flange bolts in the described criss-cross pattern.  
Tighten all flange bolts according to the described criss-cross pattern with the correct 
tightening torque. Repeat this procedure three times in total to ensure correct meter 
installation. 

Note!

 Criss-cross pattern to be repeated three times.

Summary of Contents for flowIQ 4200

Page 1: ...Installation guide flowIQ 4200 US Kamstrup Water Metering LLC 245 Hembree Park Drive Ste 110 Roswell GA 30076 USA info us kamstrup com...

Page 2: ...installation 6 4 1 Sediments in the water 7 4 2 Spool piece installation only applies to 6 and 8 meters 7 4 3 Split flange installation 8 4 3 1 General flange features and specifications 9 4 3 2 Criss...

Page 3: ...the correct flow direction Ensure correct operating pressure Avoid placing the meter in direct sunlight For meters subjected to any kind of mechanical impact it is recommended to install a lid on the...

Page 4: ...e mechanical impact For flowIQ 4200 meters it is possible to install a lid The lid can be ordered as a separate accessory and will not be attached or mounted on the meter from factory Lid type no 66 9...

Page 5: ...2 Installation precautions The sealing surface of the flanged connection must be clean and level Replace gaskets when installing a new water meter ALWAYS use new gaskets Gaskets in original quality i...

Page 6: ...er detects reverse flow The meter calculates the volume in a separate register not verified for billing Reverse volume is NOT displayed on the meter 4 flowIQ 4200 installation A flowIQ 4200 meter is d...

Page 7: ...ts to install the meter with the cup facing upwards 4 2 Spool piece installation only applies to 6 and 8 meters Spool pieces should be mounted on the pipeline flanges first to ensure flexible options...

Page 8: ...ce depends on the installation scenarios and the space conditions Both options are described below and both installation options require the same procedure for tightening the flange bolts in the descr...

Page 9: ...parts during installation Pushing nut milling If extra space for the gasket is needed insert a 7 16 nut in the pushing nut milling and insert a 7 16 bolt from the opposite side to press and enlarge th...

Page 10: ...s below 4 3 3 Installing the split flanges on the meter first Place the meter vertically on one end Fixate the two split flanges together by using a bolt Place the flange around the meter and close th...

Page 11: ...lts and nuts Tighten the bolts in the described criss cross pattern and with the correct tightening torque mentioned in the table Repeat the sequence three times to ensure that all bolts are tightened...

Page 12: ...inlets are advisable This applies to all flange meters 4 6 Flow disturbance and flow cavitation The components mentioned below can generate heavy flow disruptions or pressure drops Butterfly valve Pr...

Page 13: ...suant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate ra...

Page 14: ...been approved by Innovation Science and Economic Development Canada to operate with the antenna types listed below with the maximum permissible gain indicated Antenna types not included in this list t...

Page 15: ...riel brouilleur du Canada Cet metteur radio IC ID 22376 KWM4220 a t approuv par Innovation Sciences et D veloppement conomique Canada pour fonctionner avec les types d antenne num r s cidessous avec l...

Page 16: ...16 Kamstrup A S 55123307_A1_EN US_06 2022 Installation guide flowIQ 4200...

Reviews: