User Manual
OBM
13
When cycle has started you see this screen for manual program :
Bath temperature:
current measured bath temperature is
19.85°C
Heating power:
current heater duty cycle
14%
(from
0%
to
100%
)
Temperature status:
Stabilizing
= temperature is stabilizing and it is not yet in setpoint tolerances
Stable
= temperature is stable and it is in setpoint tolerances
Current program:
the current running program number is
0
(
0
- manual program,
1
÷
10
for automatic
program)
Setpoint temperature:
current setpoint temperature
50.00
°C
(setpoint range from
35
.00°C
to
220
.00°C
)
Stirrer speed:
the set speed for stirrer is
100%
( set range from
30%
to
100%
)
WARNING
When operating at higher temperatures set the stirring speed
at 80%!
Setpoint temperature
Heating power
Stirrer speed
Bath temperature
Current program
Temperature status