For further assistance, please contact [email protected]
3
Pulse Lite
X-Lock Amphenol (Front View)
PIN
Description
1
Digital Input 1 -
2
Digital Input 1 +
3
Digital Input 2 +
4
Digital Input 2 -
Best Practice Considerations:
•
Utilise the appropriate cable gland provided when terminating the sensor to preserve IP
rating.
o
Small: Red
– Cable OD 5.0 – 6.0mm.
o
Medium: Yellow
– Cable OD 6.0 – 8.0mm.
o
Large: Blue
– Cable OD 8.0 – 10.0mm.
•
Do not remove pins from the Amphenol plug once terminated. Please contact Kallipr for
further support
•
Ensure there is no debris between the plug and the Captis Pulse Lite prior to connecting.
1.3
Test the Sensor
When the sensor has been terminated into the provided Amphenol plug, it is recommended that the
sensor itself is tested prior to connection with the Captis device. This can be achieved by:
1. Connecting the sensor to the desired output
2. Use a multimeter (not included) on the Amphenol plug to prove the sensor output. Note that
some sensors may require power from the Captis and can only be tested once completely
connected.
Once this step is complete, the plug may be fastened to the Captis Pulse Lite.
1.4
Test Connectivity
Use the magnet provided to ‘wake’ the Captis Pulse Lite by holding it over the symbol.
The status light should flash, followed by a solid cellular light.
The device should then connect to the platform and upload data.
Prove that the sensor output is accurately recorded and uploaded by the Captis Pulse Lite.
Ensure that the device is tested in its final install location to confirm signal strength is greater than -
96dB RSRP (Reference Signal Received Power).
If the device does not connect to your platform, please refer to the Captis Device Troubleshooting
guide.
W
WAKE