Optidew User’s Manual
Kahn Instruments
33
OPERATION
3.
Particulates
– particulates passing over the sensor can build up on the
mirror over time. This can cause a loss of mirror reflectivity. DCC will
compensate for this by taking into account anything on the surface of
the mirror when resetting the optical condition, however if the problem
becomes too severe, the ‘mirror contamination warning’ symbol will be
displayed in the Sensor Status display.
Figure 26
Mirror contamination warning symbol
4.
Sample temperature
– consider the difference between the sample
temperature and the dewpoint temperature. Make sure that the sensor
you are using has the cooling capability to make the measurement (see
section 4.5. for further information). If the sensor does not have the
necessary cooling capability, then you should consider an extractive
system so the sample can be cooled prior to measurement.
5.
Sample pressure
– If you are interested in readings in terms of ppm
V
or g/m
3
Insure that the sensor is positioned in an environment of known
pressure. You can then either enter this pressure into the Optidew via the
‘Inputs’ screen (see section 3.2), or connect a pressure sensor directly to
the point of measurement (see section 2.6).
Extractive
If the sensor will be mounted into a sample conditioning system, then the above points
are still of relevance, but the following should also be considered:
1.
Extraction point
– make sure that the chosen extraction point is
representative of the process, i.e. that the sample of interest is flowing
past the extraction point, and it is not being pulled from a dead volume.
2.
Enclosure and sample line heating
– if the sample has a dewpoint
greater than ambient temperature, then all components upstream of the
sensor will need to be heated to at least 18°F/10°C above the sample
dewpoint to ensure the water remains in vapor phase.
3.
Sample block flow path
– the sensor block must be configured with
gas inlet and outlets installed in the side ports. The top is either blanked
or used to install a pressure transmitter.
If replacing an old Optidew
installation then be aware that using the old sensor block will
result in poor response speed at low dew-points, as it does not
allow enough flow across the mirror.
Summary of Contents for Optidew Series
Page 2: ...Optidew For Kahn Instruments contact information please go to www kahn com...
Page 13: ...Optidew User s Manual 4 October 2019 INSTALLATION This page is intentially left blank...
Page 53: ...Optidew User s Manual 44 October 2019 APPENDIX A Appendix A Technical Specifications...
Page 56: ...Optidew User s Manual Kahn Instruments 47 APPENDIX B Appendix B Modbus Register Map...
Page 65: ...Optidew User s Manual 56 October 2019 APPENDIX C Appendix C Dimensional Drawings...
Page 70: ...Optidew User s Manual Kahn Instruments 61 APPENDIX C Dual Stage Dewpoint Sensor...
Page 71: ...Optidew User s Manual 62 October 2019 APPENDIX C Sample Block...
Page 74: ...Optidew User s Manual Kahn Instruments 65 APPENDIX C...
Page 75: ...Optidew User s Manual 66 October 2019 APPENDIX C...
Page 76: ...Optidew User s Manual Kahn Instruments 67 APPENDIX C...
Page 77: ...Optidew User s Manual 68 October 2019 APPENDIX C...
Page 78: ...Optidew User s Manual Kahn Instruments 69 APPENDIX C...
Page 79: ...Optidew User s Manual 70 October 2019 APPENDIX C...
Page 80: ...Optidew User s Manual Kahn Instruments 71 APPENDIX C...
Page 81: ...Optidew User s Manual 72 October 2019 APPENDIX C...
Page 82: ...Optidew User s Manual Kahn Instruments 73 APPENDIX C...
Page 83: ...Optidew User s Manual 74 October 2019 APPENDIX C...
Page 84: ...Optidew User s Manual Kahn Instruments 75 APPENDIX C...
Page 85: ...Optidew User s Manual 76 October 2019 APPENDIX C...
Page 86: ...Optidew User s Manual Kahn Instruments 77 APPENDIX C...
Page 87: ...Optidew User s Manual 78 October 2019 APPENDIX C...
Page 88: ...Optidew User s Manual Kahn Instruments 79 APPENDIX C...