Optidew User’s Manual
14
October 2019
OPERATION
2.4.3 Integral Sensor mounting
If the sensor is to be positioned into a sealed but open environment (glove box,
environmental chamber or area to be monitored) a female thread of M36 x 1.5-6H is
required to suitably thread onto the sensors mail M36 x 1.5-6g thread. The bonded
seal provided will require a good surface finish (0.8 Ra) across a minimum sealing face
of DIA 46.0mm to insure leak free operation up to the max operating pressure of the
sensor (25barg). The bonded seal will also require a strong hand tightening to insure
leak free sealing of the two mating faces.
Insure that the sensor is suitably secured to prevent any movement and that it is
located in a position that will see a representative flow of the sample to be measured.
2.5
Temperature Probe Installation
The temperature probe is supplied pre-wired and simply requires fitting to the connector
on the Optidew control unit prior to use.
Take into consideration how you will use the readings from your temperature probe
before installing it. If the measurement will be used in combination with the dewpoint
measurement to calculate % RH, then the temperature probe should be installed in
a location which is most representative of the temperature of your environment or
sample.
Be aware that when depressing the mirror temperature by more than 40°F the Optidew
dewpoint sensor will generate a small amount of heat in the surrounding area. Try to
situate the temperature probe upstream of the dewpoint sensor and at least 6 inches
away.
Refer to section 4.4.5 for more information on calculated parameters and which
measured inputs are used to derive them.
2.6
Pressure Transmitter Installation
Pressure transmitters are available for the Optidew in several ranges. Any 4-20mA
pressure transmitter can be wired into the Optidew control unit via the 4-pin M12
connector.
Kahn can supply a pressure transmitter with the Optidew, which is installed via a 1/8”
NPT male thread.
For ease of use, the dewpoint sensor and pressure transmitter should be installed so
they are operating at the same pressure; i.e. either both at atmospheric pressure or
both at line pressure.
For information on the pressure compensation feature refer to section 4.4.5.
Summary of Contents for Optidew Series
Page 2: ...Optidew For Kahn Instruments contact information please go to www kahn com...
Page 13: ...Optidew User s Manual 4 October 2019 INSTALLATION This page is intentially left blank...
Page 53: ...Optidew User s Manual 44 October 2019 APPENDIX A Appendix A Technical Specifications...
Page 56: ...Optidew User s Manual Kahn Instruments 47 APPENDIX B Appendix B Modbus Register Map...
Page 65: ...Optidew User s Manual 56 October 2019 APPENDIX C Appendix C Dimensional Drawings...
Page 70: ...Optidew User s Manual Kahn Instruments 61 APPENDIX C Dual Stage Dewpoint Sensor...
Page 71: ...Optidew User s Manual 62 October 2019 APPENDIX C Sample Block...
Page 74: ...Optidew User s Manual Kahn Instruments 65 APPENDIX C...
Page 75: ...Optidew User s Manual 66 October 2019 APPENDIX C...
Page 76: ...Optidew User s Manual Kahn Instruments 67 APPENDIX C...
Page 77: ...Optidew User s Manual 68 October 2019 APPENDIX C...
Page 78: ...Optidew User s Manual Kahn Instruments 69 APPENDIX C...
Page 79: ...Optidew User s Manual 70 October 2019 APPENDIX C...
Page 80: ...Optidew User s Manual Kahn Instruments 71 APPENDIX C...
Page 81: ...Optidew User s Manual 72 October 2019 APPENDIX C...
Page 82: ...Optidew User s Manual Kahn Instruments 73 APPENDIX C...
Page 83: ...Optidew User s Manual 74 October 2019 APPENDIX C...
Page 84: ...Optidew User s Manual Kahn Instruments 75 APPENDIX C...
Page 85: ...Optidew User s Manual 76 October 2019 APPENDIX C...
Page 86: ...Optidew User s Manual Kahn Instruments 77 APPENDIX C...
Page 87: ...Optidew User s Manual 78 October 2019 APPENDIX C...
Page 88: ...Optidew User s Manual Kahn Instruments 79 APPENDIX C...