Easidew I.S. User’s Manual
Kahn Instruments
3
INSTALLATION
2.2
Preparation of the Sensor Cable
The sensor cable is NOT supplied as standard. A cable can be obtained by contacting
Kahn Instruments (see www.kahn.com for details).
The crimps supplied must be fitted onto any cable installed
into the connector in order to comply with Hazardous Area
Certification of the product.
If making a cable assembly it is important that the cable is correctly terminated. See
Figures 3 to 6.
Cable connection to the Easidew I.S. transmitter is made via the removable connector.
Removing the central screw enables the connector terminal block to be removed from
the outer housing by using a small screwdriver to pry it off.
O-ring
and washer
Figure 2
Connector Terminal Block Removal
Caution: When removing the central screw insure that the small
sealing O-ring and the washer are retained on the screw and
are present during re-installation.
NOTE:
Figure 3
to
Figure 6
shown below, should be followed in detail. The
crimps should be applied such that there is no possibility of a conductor
strand of a core becoming free (see
Figure 4).
Figure 3
Bare Wires
Figure 4
Crimped Wires