Easidew I.S. User’s Manual
97099 Issue 13.2, April 2015
21
APPENDIX C
Appendix C
Hazardous Area Certification
The Easidew I.S is certified compliant to the ATEX Directive (94/9/EC), and IECEx for
use within Zone 0, 1 and 2 hazardous areas and has been assessed so by Baseefa Ltd
(Notified Body 1180).
The Easidew I.S is certified compliant to the North American Standards (USA and
Canada) for use within Class I, Division 1, Groups A, B, C & D hazardous locations and
has been assessed so by CSA and FM.
C.1
Product Standards
This product conforms to the Standards:
EN60079-0:2012
IEC60079-0:2011
EN60079-11:2012
IEC60079-11:2011
FM Class 3600:1998
CAN/CSA-C22.2 No. 0-10
FM Class 3610:2007
CAN/CSA-C22.2 No. 157-92
FM Class 3810:2005
C22.2 No. 142-M1987
C.2
Product Certification
This product is attributed with the product certification codes:
ATEX & IECEx
II 1G Ex ia IIC T4 Ga (-20°C to +70°C)
North American
IS, Class I, Division 1, Groups A, B, C & D, T4
C.3
Global Certificates/Approvals
ATEX
Baseefa 06ATEX0330X
ATEX System Baseefa 07Y0027
IECEx
IECEx BAS 06.0090X
CSA
2013218
FM
3030238
TC TR Ex
RU C-GB. ГБ05.B.00229
These certificates can be viewed or downloaded from our website at:
http://www.kahn.com
C.4
Terminal Parameters
Ui
= 28 V
li
= 93 mA
Pi
= 651 mW
Ci
= 37 nF
Li
= 0