Dewert User Manual Heavy-Weight Tables – Updated 11.2021/3 MDR                                                         Page 2 of 14 

1. Safety information 

 
1.1 Applied 

symbols 

Safety instructions and key sections in this user manual are marked with the exclamation mark symbol on the 
left. Please pay particular attention to these instructions and sections. 
 
Other symbols possibly used on the device: 

Observe instructions for use: 

 

 

Risk of pinching when adjusting:  

 

Risk of hazard zone: 

 

 

 

 

 

 

1.2 Applied standards 

 

This device has been designed and manufactured in accordance with national and international regulations. 
This ensures a very high level of equipment safety. 
The models described here comply with the following regulations and directives: 
 

Regulation (EU) 2017/745 
DIN EN IEC 60601-1 
DIN EN ISO 14971 
DIN EN 60601-2-52 partly based on 
DIN EN IEC 62353 
DIN EN ISO 10993-5/-10 
DGUV Regulation 3 

     
This device is a Class 1 medical device according to Regulation (EU) 2017/745 (MDR). 
 
 

1.3 Safety regulations

 

This section contains a summary of the most important safety information. 
 
Correct operation of the device is essential for safe operation. Therefore, please familiarise yourself with the 
contents of these instructions for use before using the device. We recommend that you keep these 
instructions for use near the device for future reference.  
The device may only be used by authorised, instructed and competent persons who are sufficiently familiar 
with its adjustment mechanism or have read and understood the operating manual fully. The manufacturer 
cannot be held responsible for damage caused by or involving unauthorised persons. No third-party devices 
may be installed without consultation with us or brought in the direction of movement of the table in such a 
way that a possible hazard potential arises. 

 

The user must ensure that the device is not accessible to unauthorised persons or cannot be operated by 
unauthorised persons even when left unattended. 
When leaving the device, it should be secured in such a way that unauthorised adjustment is impossible. 

Important:

 Never leave the device unattended or accessible to third parties when it is ready for operation. 

 

Keep a sufficient safety distance to the device during all adjustment procedures. Special attention 
must be paid to the arms, hands, legs and feet of the user and the patient - 

RISK OF CRUSHING!

 

 

 

Make sure that there are no objects located directly around or underneath the device! 

 

 

 

 

 

1.4 Intended purpose

 

The table is used for the ideal positioning of patients for the purpose of curative and disease treatment, 
examination, massage and health therapy. 

Summary of Contents for 208

Page 1: ...mbols 2 1 2 Applied standards 2 1 3 Safety regulations 2 1 4 Intended purpose 2 1 5 Information on setup and use 3 1 6 Commissioning 3 1 7 Safety instructions 3 1 8 Model designation and type labelling 4 1 9 Meaning of the serial number 4 2 Operating manual 5 2 1 Table design 5 2 2 Height adjustment 5 2 3 Head part adjustment 5 2 4 Movability model dependent or optional 5 2 5 Additional equipment ...

Page 2: ...the device We recommend that you keep these instructions for use near the device for future reference The device may only be used by authorised instructed and competent persons who are sufficiently familiar with its adjustment mechanism or have read and understood the operating manual fully The manufacturer cannot be held responsible for damage caused by or involving unauthorised persons No third ...

Page 3: ...e do not lift it at the head part as this may damage the head part and or the release mechanism Hold and lift the table on the left and right hand side of the frame The table must stand securely on its feet or castors on a level flat and solid surface Before use activate the brakes on the castors and make sure they are working properly When transporting the table take hold of the underframe NOT th...

Page 4: ...during maintenance work To disconnect be sure to grasp the plug not the power cable This device must not be modified without the express permission of the manufacturer When transporting the table only take hold of the underframe NOT the upper frame 1 8 Model designation and type labelling The exact model designation depends on the choice of frame colour 00 white powder coated RAL 9010 03 white alu...

Page 5: ...ke place via the foot switch or the hand switch Please refer to section 7 The table is lifted or lowered Note on operation The electric motor is to be operated in intermittent duty mode This means that a maximum duty cycle of 25 s must not be exceeded Before switching the motor back on an interruption time of at least 400 s must be observed If the maximum duty cycle is exceeded an internal thermal...

Page 6: ... second hole Then relieve the spring tension on the sleeve To release the paper roll holder proceed in the same way Push handles optional Depending on the actual equipment push handles are located either at the end of the head part or the foot part or on both sides When moving the table always use both hands on the push handles to push pull the table Side guard optional Foldable side guard Folding...

Page 7: ...impact protection Additional foot switch Additional hand switch Foot switch fixation on the underframe 4 Technical data 4 1 Individual models 208 209 7209E 7255E Models Max length mm 1950 1950 1950 1990 Width mm 650 800 800 800 Head part length mm 540 540 500 540 Weight approx depending on equipment kg 60 70 110 130 Min max height mm 540 890 590 930 Fixed height mm 650 650 Adjustment time height m...

Page 8: ...n the meaning of EN 50082 Part 1 and 2 5 Cleaning instructions Upholstery covering There are two different collections of upholstery covering to choose from Skai Pandoria Plus manufactured by Hornschuch Continental in Germany Skai Toronto EN manufactured by Hornschuch Continental in Germany For cleaning and disinfection a selection of agents from various manufacturers has been tested for compatibi...

Page 9: ...wivel joints for completeness Check proper fit of the screw connections Lubricate the swivel joints and operating levers slightly with low viscosity spray oil if necessary Lubricate the roller guides with a little bearing grease or similar using a brush if necessary Insufficient lubrication is noticeable through noise development A checklist for maintenance technical inspection can be found in the...

Page 10: ...the control unit first press the desired direction of travel for approx 1 second on the control element After a short waiting time 1 2 seconds press the desired direction of travel again and the actuator can be moved for a maximum of 30 seconds in this direction of travel If the switching cycle of the double tap function is not observed the actuator cannot be operated After 30 seconds of operation...

Page 11: ...sal Packaging Safety notice Pay attention to sharp edges and pointed objects during disposal The packaging materials produced are mainly o Cardboard Paper o Plastic o Wood when delivered on a pallet Please observe the local regulations for waste disposal and preferably recycle the materials As a manufacturer we are licensed as a participant in the Dual System in accordance with the German Packagin...

Page 12: ...s use or who through experience with other similar medical devices have knowledge of its proper use taking into account possible hazards Equipped with the options of movability not wheel lifting fixing mechanism side guards and push handle the intended purpose of the table is extended and also provides for the ideal positioning of patients for the purpose of transport to pre treatment or post trea...

Page 13: ...d supply lines undamaged Function check With electromotive adjustment Move all motors with the foot switch or the hand switch to both limit positions until automatic switch off to ensure that the table operates smoothly without any collisions or blockages no cable connection can be overstretched crushed or otherwise damaged the motors run without noticeable noise development the end position switc...

Page 14: ...le All necessary toggle screws present Possible movability Castors undamaged freewheeling given Connection to frame undamaged Re tighten all screw connections with central locking system also the grub screws of switch levers Safe braking effect Check with locked brakes by pulling and pushing the table Central Movability Levers must be in contact with the frame tighten the fastening screws Shift le...