Always operate the side guard with the necessary caution.
Never operate the side guard if the hands, fingers, etc. of another person are located between the
bars or on the mechanism of the side guard. Risk of being crushed/pinched!!!
The laterally lowerable side guard is completely screwed in place. If the side guard becomes too loose or has
excessive lateral play over time due to use, it can be readjusted by tightening the screws. The moving parts of
the side guard should be re-lubricated slightly at regular intervals (spray oil, e.g. Ballistol).
3. Additional accessories (for user-specific table configuration), in extracts
•
Heavy-weight castors Ø 125 mm, centrally lockable
•
Paper roll holder
•
DIN rails
•
Fastening clamp for DIN rail
•
Infusion pole for fastening clamp
•
Armrest for DIN rail and fastening clamp
•
Push handle
•
Nose opening in the head part and cushion
•
Electromotive head part adjustment
•
Wall protection bumpers / impact protection
•
Additional foot switch
•
Additional hand switch
•
Foot switch fixation on the underframe
4. Technical data
4.1 Individual models
208
209
7209E
7255E
Models
Max. length (mm)
1950
1950
1950
1990
Width (mm)
650
800
800
800
Head part length (mm)
540
540
500
540
Weight (approx., depending on
equipment) kg
60
70
110
130
Min.
– max. height (mm)
/
/
540
– 890
590
– 930
Fixed height (mm)
650
650
/
/
Adjustment time, height (motor) (s)
/
/
36
36
Head part adjustment range (gas spring)
0° / +40°
0° / +40°
0° / +40°
/
Head part adjustment range (motor)
Optional
Optional
Optional
0° / +55°
Foot part adjustment range (motor)
/
/
/
0° / +30°
Max. patient weight (kg)
400
400
400
350
The motors are equipped with a thermal protective switch as standard. This causes the motors to cut out if the duty
cycle (DC: 25 s / 400 s) is exceeded. The table should be ready for use again after a 15-minute rest period. There
is, therefore, no risk of overloading the motor. The engine has been specially designed for this high weight load.
The mechanics of the tables have been designed with extensive safety margins in mind. The max. patient weight is
provided with a 4-fold static safety factor, i.e. the design has been fully tested for a 4-fold load.
Technical data subject to change