- 19 -
Reference
Power save function
The monitor enters into power save mode after idling for a certain period of time to reduce
power consumption. Refer to the computer instruction manual on how to set it.
Digital input
This monitor supports only DVI digital signal. It is normal for noise to occur during
startup, shutdown, and/or at switching resolutions. This is caused by the graphics card in
the system as it switches resolution. Also, when switching resolutions, a 5 sq. mm white
area may appear under the front sensor. In both cases, they are not failures.
Note
Turn off the system when connecting and disconnecting the DVI cable. It could
damage the graphics card.
Operating Principle
Image signals from acquisition devices or picture-archiving communication system (PC)
shall be transmitted to the monitor via DVI or DisplayPort terminals.
By transmitting image signals of the system to the input unit of the monitor, the monitor
can provide images on its screen (LCD panel) via its signal processing circuit. It is
possible for users to configure the monitor freely for suitable observation by OSD
because of the input unit and the signal processing circuit which correspond to various
input signals.
Input unit
Signal
processing
circuit
Power unit
Operation unit
LCD Module