Guidance and manufacturer’s declaration – electromagnetic immunity
The model CL-S200 is intended for use in the electromagnetic environment specified below.
The customer or the user of the model CL-S200 should assure that it is used in such an
environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment
–guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines at 100 kHz rate
±2 kV for power
supply lines
±1 kV for input/output
lines at 100 kHz rate
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
0 %
U
T
; 0,5 cycle
At 0, 45, 90, 135, 180,
225, 270, 315
°
0 %
U
T
;
1 cycle and 70
%
U
T
: 25/30 cycles
At 0
°
0 %
U
T
; 250/300 cycles
0 %
U
T
; 0,5 cycle
At 0, 45, 90, 135, 180,
225, 270, 315
°
0 %
U
T
;
1 cycle and 70
%
U
T
: 25/30 cycles
At 0
°
0 %
U
T
; 250/300 cycles
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the model
CL-S200 requires
continued operation
during power mains
interruptions, it is
recommended that the
model CL-S200 be
powered from an
uninterruptible power
supply or a battery.
Power
frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.