First, select between 12h mode 24h mode; then set date and time. 
Settings  for  date  and  time  are  not  erased  when  the  device  is 
power off. 

 

3)Language 
 
 

 

 
 
 
Depending on the firmware, up to two different language options 
are available here for selection. All messages and menus will be 
displayed in the selected language. 
 
 
 
 
 
 
 

Service 

The Service submenu is protected by a PIN code; only authorized 
service personnel can access this menu. 

 When the ambient temperature is 20

, the time required for 

the monitor to warm from the minimun storage temperature 
between uses until it is ready for its intended use is 1 hour.

 

 

When the ambient temperature is 20

, the time required for 

the  monitor  to  cool  from  the  maximun  storage  temperature 
between uses until it is ready for its intended use is 4 hours. 

5.Erro Messages –Problems- Corrective Actions

 

5.1 General Information

 

Physiological conditions, medical procedures, or external agents 
that may interfere with the monitor’s  ability to detect and display 
accurate measurements include: 

·

Lint,dust,light(including sunlight) 

·

Incorrect applications of  the sensor 

·

Degraded sensors and electrodes, or loosened electrodes 

·

Placement of the senor on an extremity with a blood pressure 
cuff, arterial catheter, or intravascular line 

·

Excessive patient activity 

·

Intravascular dyes 

·

Externally applied colouring agents, such as nail polish 

·

Failure to cover the sensor site with opaque material in high 
ambient light conditions. 

·

Venous pulsation 

·

Dysfunctional haemoglobin 

·

Low perfusion 

Electromagnetic  interference  may  be  preventing  the  monitor 

5.2 Error Messages- Causes 

No sensor

 

The sensor is not connected properly to the device. 
–  Check sensor connection 
 

Probe off!

 

The sensor has been removed from the monitoring site. –  Check 
that the sensor is properly attached to the patient. 
 

Low battery!

, battery symbol blinking red 

The battery is almost completely discharged. 
-Replace batteries immediately. 
 

Sensor fault

 

The connected sensor is either defective or not compatible with the 
device – check sensor. 
 

Device defective

 

Fatal device error, e.g. resulting from improper handling, such as 
use with computed tomography. – The device must be sent in to the 
Service Department. 
 

Too much ambient light

!

 

High ambient light sources near the sensor, e.g. surgical lights.  
–  Shield sensor more effectively from external light. 
 

Bad signal quality

 

Poor-quality pulse signal, for example as a result of low perfusion. 
– Move the sensor to a different site on the patient or provide more 
effective monitoring conditions. 
 

5.3 Failure – Cause – Corrective Action 
 
Problem: There is no response to the Power button.  

Cause – Corrective Action: Ensure that the Power button is fully 
depressed.  
 

Problem: No pulse signal found 

Cause – Corrective Action: Check the patient. Check the sensor 
directions for use to determine if an appropriate sensor is being 
used and if it is applied properly. 
 
Check sensor and extension cable connections. Test the sensor on 
another subject. Try another sensor or extension cable. 

 

Perfusion may be too low for the monitor to track the pulse. Check 
the patient. Test the monitor on yourself. Change the sensor site. 
Try another sensor. 
 
Interference due to patient activity may be preventing the monitor 
from tracking the pulse. 
 
Keep the patient still, if possible. Verify that the sensor is securely 
applied and replace it if necessary. Change the sensor site. 
 
The sensor may be too tight, there may be interference due to 
ambient light, or the sensor may be on extremity with a blood 
pressure cuff, arterial catheter, or intravascular line. Reposition 
sensor, as necessary. 

Setup 

 

Data transmission 
Date  and Time 
Language

 

 

Date and Time 

 

Data : 2015  / 01 / 01 
Time:06:08 
Exit

 

 

 

Setup 

 

Data transmission 
Date  and Time 
Language

 

 

Language 

 

English 
Chinese 
Exit

 

Setup 

 

Language 
Service 
Exit

 

 

Service 

PIN Code: 

0   0   0   0 

Enter 
Exit 

 

Summary of Contents for OxiEasy W60

Page 1: ...JT OxiEW60 UM01 OxiEasy W60 Pulse Oximeter User Manual Easy Reliable and Smart...

Page 2: ...rnings 3 2 Parts Instruction 4 3 Operation instructions 5 4 Screen off Device 5 5 Erro Messages Problems Corrective Actions 6 6 Battery and Charger 7 7 Maintenance Cleaning 7 8 Specification 8 9 Testi...

Page 3: ...lid reading of the W60 Failure to cover the sensor site with opaque material in high ambiernt light conditions may result in inaccurate measurement Disconnect the OxiEasy W60 and sensor from the patie...

Page 4: ...ry level The icon will flash and turn red if the battery is low 2 Low Bat Low voltage alarm 3 98 SpO 2 The SpO2 value shows the blood oxygen saturation level expressed as a percentage The small number...

Page 5: ...o to turn on or turn off the transmission 2 Date and time 4 Screen of Device 4 1 Main Menu All important and frequently used setting are accessible through the main menu which can be opened by pressin...

Page 6: ...ly attached to the patient Low battery battery symbol blinking red The battery is almost completely discharged Replace batteries immediately Sensor fault The connected sensor is either defective or no...

Page 7: ...lular phones Mobile two way radios Electrical appliances High definition televisions HDTV s The monitor is designed for use in environments in which the pulse can be obscured by electromagnetic interf...

Page 8: ...nes whether the oximeter meets the criteria of ISO 80601 2 61 for pulse rate during simulated movement tremor and spike motions 10 Warranty We give 1 year warranty for defects caused by material or ma...

Page 9: ...cement Parts 12 1 Packing List OxiEasy W60 main unit Sensor USB Connector CD 12 2 Main Accessories Description Main unit USB Connector Sensor Model Manufacturers Structure S18 S8137 Himed S16 S7137 Hi...

Page 10: ...60 User Manual 10 JUSTEC Justec Shenzhen CO LTD Add Room 2018 2nd Floor 64th Building Jinlong Industrial Area Qilin Road Nanshan District 518052 Shenzhen People s Republic of China Tel 0086 755 260689...

Reviews: