 OxiEasy W60 User Manual                                                                                                                                                               7 

from tracking the pulse. Remove the source of interference. 
 

Problem: After a valid measurement the pulse signal can not 
be found anymore 

Cause  – Corrective Action: Check the patient. Check the sensor 
directions  for  use  to  determine  if  an  appropriate  sensor  is  being 
used  and  if  it  is  applied  properly.  Check  sensor  and  extension 
cable  connections.  Test  the  sensor  on  another  subject.  Try 
another sensor or extension cable. 
 
Perfusion  may  be  too  low  for  the  monitor  to  track  the  pulse. 
Check  the  patient.  Test  the  monitor  on  yourself.  Change  the 
sensor site. Try another sensor. 
 
Interference  due  to  patient  activity  may  be  preventing  the 
monitor from tracking the pulse. 
 
Keep  the  patient  still,  if  possible.  Verify  that  the  sensor  is 
securely  applied  and  replace  it  if  necessary.  Change  the  sensor 
site. 
 
The  sensor  may  be  too  tight,  there  may  be  interference  due  to 
ambient light, or the sensor may be on an extremity with a blood 
pressure  cuff,  arterial  catheter,  or  intravascular  line.  Reposition 
sensor, as necessary. 
 
Electromagnetic  interference  may  be  preventing  the  monitor 
from tracking the pulse. Remove the source of interference. 
 

Other problems: 

EMI (Electromagnetic Interference) 
Caution:  This  device  has  been  tested  and  found  to  comply  with 
the  limits  for  medical  devices  according  to  EN  60601-1-2, 
(second  edition),  and  the  Medical  Device  Directive  93/94/EEC. 
These  limits  are  designed  to  provide  reasonable  protection 
against harmful interference in a typical medical installation. 
 
Due  to  the  proliferation  of  radio-frequency  transmitting 
equipment  and  other  sources  of  electrical  noise  in  healthcare 
environments, it  is possible that high levels of such interference 
due  to  close  proximity,  or  strength  of  a  source,  may  result  in 
disruption  of  performance  of  this  device.  Examples  of  noise 
sources  in  healthcare  environments  and  home  care  environment 
that could cause electromagnetic interference are: 

 

·

Electrosurgical units 

·

Cellular phones 

·

Mobile two-way radios 

·

Electrical appliances 

·

High-definition televisions(HDTV

s) 

 
The  monitor  is  designed  for  use  in  environments  in  which  the 
pulse  can  be  obscured  by  electromagnetic  interference.  During 
such  interference,  measurements  may  seem  inappropriate  or  the 
monitor may not seem to operate correctly. 
Disruption  may  be  evidenced  by  erratic  readings,  cessation  of 
operation, or other incurred functioning. If this occurs, the site of  
3) Clean up the machine carefully with a soft tower. 

 

Use should be surveyed to determine the source of disruption, and 
the following actions taken to eliminate the source. 
 

·

Turn equipment in the vicinity off and on to isolate the offending 
equipment. 

·

Reorient or relocate the interfering equipment. 

·

Increase the distance between the interfering equipment and this 
equipment. 

 
The monitor generates, uses, and radiates radio frequency energy. If 
it is not installed and used in accordance with these instructions. 
The monitor may cause harmful interference with other devices in 
the vicinity. 

6.Battery and Charger 

 

6.1 Battery 

The OxiEasy W60 use 3.7V rechargeable lithium battery. 
 

6.2 Charging method 

·

The metal contacts on the bottom of the instrument are connected 
to the probes on the charger, See below picture: 

 

             

 

·

Put the usb connector of the charger into the power adaptor, 

    the indicator light will be green. 

·

During the charging process, the indicator of the screen can show 
the charging process. When battery is full

the signal of the 

power indicator  will stop beating. 

Please use specified charger and recommended charge methods 

 Don’t expose the battery in high or cold temperature, otherwise 

it will affect the battery life. When charging, the current will 
have a return, then the battery is with a little heat, which is 
normal. 

7.Maintenance – Cleaning 

The device requires no calibration .  
The  device

s  expected  service  life  is  5  years.  The  sensor

expected service life is 3 years. 

 

7.1 Maintenance 

The device should be examined every 12 months as bellow: 
1) Check if there are mechanical or function problem of device. 
2) Check the label on the device and make sure it is clear. 
We give no warranty if the operator has destroyed the device by 

Summary of Contents for OxiEasy W60

Page 1: ...JT OxiEW60 UM01 OxiEasy W60 Pulse Oximeter User Manual Easy Reliable and Smart...

Page 2: ...rnings 3 2 Parts Instruction 4 3 Operation instructions 5 4 Screen off Device 5 5 Erro Messages Problems Corrective Actions 6 6 Battery and Charger 7 7 Maintenance Cleaning 7 8 Specification 8 9 Testi...

Page 3: ...lid reading of the W60 Failure to cover the sensor site with opaque material in high ambiernt light conditions may result in inaccurate measurement Disconnect the OxiEasy W60 and sensor from the patie...

Page 4: ...ry level The icon will flash and turn red if the battery is low 2 Low Bat Low voltage alarm 3 98 SpO 2 The SpO2 value shows the blood oxygen saturation level expressed as a percentage The small number...

Page 5: ...o to turn on or turn off the transmission 2 Date and time 4 Screen of Device 4 1 Main Menu All important and frequently used setting are accessible through the main menu which can be opened by pressin...

Page 6: ...ly attached to the patient Low battery battery symbol blinking red The battery is almost completely discharged Replace batteries immediately Sensor fault The connected sensor is either defective or no...

Page 7: ...lular phones Mobile two way radios Electrical appliances High definition televisions HDTV s The monitor is designed for use in environments in which the pulse can be obscured by electromagnetic interf...

Page 8: ...nes whether the oximeter meets the criteria of ISO 80601 2 61 for pulse rate during simulated movement tremor and spike motions 10 Warranty We give 1 year warranty for defects caused by material or ma...

Page 9: ...cement Parts 12 1 Packing List OxiEasy W60 main unit Sensor USB Connector CD 12 2 Main Accessories Description Main unit USB Connector Sensor Model Manufacturers Structure S18 S8137 Himed S16 S7137 Hi...

Page 10: ...60 User Manual 10 JUSTEC Justec Shenzhen CO LTD Add Room 2018 2nd Floor 64th Building Jinlong Industrial Area Qilin Road Nanshan District 518052 Shenzhen People s Republic of China Tel 0086 755 260689...

Reviews: