1. Intended Use 

 
The OxiEasy W60 pulse oximeter is intended for  spot-checking 
and continuous monitoring arterial oxygen saturation (SpO

2

) and 

pulse rate of adult and pediatric patients in hospital, hospital type 
facilities, transport, emergency care, mobile environments and 
home care. 
 

Warnings 

 
Warnings  are  indentified  by  the  WARNING  symbol  shown 
above.  Warnings  alert    the  user  to  potential  serious  outcomes, 
such as death, injury, or adverse events to the patient or user. 
 

 

Please read the instructions carefully to ensure accurate 

measurement and safe operation.

 

 

 Do  not  make  any  clinical  judgments  based  solely  on  the 

OxiEasy W60. The monitor is intended only as an adjunct in 
patient  assessment.  It  must  be  used  in  conjunction  with 
clinical  signs  and  symptoms.  The  interpretation  of  the 
measurement  values  should  be  done  only  by  trained  health 
care professionals.  

 

 Explosion hazard. Do not use OxiEasy W60 in the presence 

of  flammable  anesthetics  mixture  with  air,  oxygen,  or 
nitrous oxide. 

 

 Routinely  monitor  the  patient  to  make  sure  the  OxiEasy 

W60 is functioning and that the sensor is correctly placed. 

 

Pulse  oximetry  measurements  and  pulse  signals  can  be 

affected  by  certain  environmental  conditions,  sensor 
application  errors,  and  certain  patient  conditions.  See  the 
appropriate  sections  of  this  manual  for  specific  safety 
information. 

 If you are uncertain about the accuracy of any measurement, 

check the patient's vital signs by alternate means; then make 
sure that the OxiEasy W60 is functioning correctly. 

 

 The  use  of accessories ,sensors, and  cables other  than those 

specified  may  result  in  inreased  emission  and/or  create 
invalid reading of the W60. 

 

 Failure to cover the sensor site with opaque material in high 

ambiernt  light  conditions  may  result  in  inaccurate 
measurement. 

 

 

 Disconnect  the  OxiEasy  W60  and  sensor  from  the  patient 

throughout  magnetic  resonance  imaging(MRI)  scanning. 
Induced could potentially cause burns. 

 
 

 

The  OxiEasy  is  not  defibrillator-proof.  However,  it  may 

remain  attached  to  the  patient  throughout  defibrillation  or 
whilst  an  electrosurgical  unit  is  in  use.  The  measurements 
may be inaccurate throughout the defibrillation, or use of an 
electrosurgical unit, and shortly  thereafter. To  avoid shock, 
the  caregiver  should  not  touch  the  OxiEasy  while  using  a 
defibrillation on a patient.

 

 

The  OxiEasy  W60  is  a  prescription  device  to  be  operated 

only  by  trained  personnel.  The  monitor  is  for  attended 
monitoring only.

 

 

 Do  not  use  a  device  that  appears  damaged.  Do  not  use  the 

sensors where optical components lie open. 

 

 Changing  lithium  batteries  when  incorrect  replacement  by 

inadequately  trained  personnel  would  result  in  an 
unacceptable  risk(e.g.,  excessive  temperatures,  fire  or 
explosion)

 

 

 

Do not modify this equipment without authorization of the 

manufacturer. 

 

 Check the pulse oximeter sensor application site every 4 

hours to determine the positioning of the sensor and the 
circulation and skin sensitivity of the patient. Patient 
sensitivity varies depending on medical status or skin 
condition.

 

 

 

Carefully route patient cables and connections to reduce the 

possibility of patient entanglement, strangulation, or 
injury to the patient.

 

 

 

When using the monitor around pets,small child or baby, 

avoid leaving the monitor unattended. Cables pose a 
risk of injury, including strangulation. This device contains 
small parts that may cause suffocation if swallowed by child 
or baby.

 

 

 

Do not use the device when alarms are required.

 

 

  

When using the monitor in the home, avoid exposing the 

monitor to lint and dust.

 

 

 

 

 

 

 

 
 

Summary of Contents for OxiEasy W60

Page 1: ...JT OxiEW60 UM01 OxiEasy W60 Pulse Oximeter User Manual Easy Reliable and Smart...

Page 2: ...rnings 3 2 Parts Instruction 4 3 Operation instructions 5 4 Screen off Device 5 5 Erro Messages Problems Corrective Actions 6 6 Battery and Charger 7 7 Maintenance Cleaning 7 8 Specification 8 9 Testi...

Page 3: ...lid reading of the W60 Failure to cover the sensor site with opaque material in high ambiernt light conditions may result in inaccurate measurement Disconnect the OxiEasy W60 and sensor from the patie...

Page 4: ...ry level The icon will flash and turn red if the battery is low 2 Low Bat Low voltage alarm 3 98 SpO 2 The SpO2 value shows the blood oxygen saturation level expressed as a percentage The small number...

Page 5: ...o to turn on or turn off the transmission 2 Date and time 4 Screen of Device 4 1 Main Menu All important and frequently used setting are accessible through the main menu which can be opened by pressin...

Page 6: ...ly attached to the patient Low battery battery symbol blinking red The battery is almost completely discharged Replace batteries immediately Sensor fault The connected sensor is either defective or no...

Page 7: ...lular phones Mobile two way radios Electrical appliances High definition televisions HDTV s The monitor is designed for use in environments in which the pulse can be obscured by electromagnetic interf...

Page 8: ...nes whether the oximeter meets the criteria of ISO 80601 2 61 for pulse rate during simulated movement tremor and spike motions 10 Warranty We give 1 year warranty for defects caused by material or ma...

Page 9: ...cement Parts 12 1 Packing List OxiEasy W60 main unit Sensor USB Connector CD 12 2 Main Accessories Description Main unit USB Connector Sensor Model Manufacturers Structure S18 S8137 Himed S16 S7137 Hi...

Page 10: ...60 User Manual 10 JUSTEC Justec Shenzhen CO LTD Add Room 2018 2nd Floor 64th Building Jinlong Industrial Area Qilin Road Nanshan District 518052 Shenzhen People s Republic of China Tel 0086 755 260689...

Reviews: